8 Manual Control

tiamo 2.2 (ProcessLab)

■■■■■■■■

1511

730, 774, 778, 789
Range

0.01 ... 160.00 mL/min

Titrino
Range

0.01 ... 150.00 mL/min

Filling rate

The buret is filled at this rate after the dosing. The maximum filling rate
depends on the cylinder volume of the exchange unit or dosing unit used.
If the entered filling rate is too high for the selected dosing device, it will
automatically be reduced during dosing to the largest possible value.

Note

For viscous liquids the filling rate should be reduced.

Selection

maximum

Default value

maximum

Titrando, 814, 815, 855
Range

0.01 ... 166.00 mL/min

730, 774, 778, 789
Range

0.01 ... 160.00 mL/min

Titrino
Range

0.01 ... 150.00 mL/min

Fill automatically

on | off (Default value: on)

This parameter is only visible for devices of the type Titrando, Dosing Inter-
face and USB Sample Processor.

If this option is enabled, the buret is filled automatically after dosing. During
the filling procedure the volume display is reset to 0.0000 mL. If this option
is disabled, the added volume is displayed cumulatively.

[Start]

Start fixed volume dosing for the selected dosing device. The dosed volume
is displayed live.