3 Workplace

tiamo 2.2 (ProcessLab)

■■■■■■■■

131

3 - Start test
The following checks and actions are triggered with start test:

Authorization for carrying out the method?
Check whether the registered user is authorized to carry out the method
required (see Chapter 5.4.2.3, page 410).

Method executable?
Check whether the required method is executable. For methods which
are not executable, a prompt appears to perform the method check and
to modify the method accordingly (see Chapter 5.2.5, page 394).

Check device (modules) and rack
Check whether the instruments (and/or instrument types), device mod-
ules (e.g. measuring input, tower, etc.) and racks required in the instru-
ment-specific commands are present and available and/or assigned. If
the instrument in a command is not yet assigned, then the device
assignment
dialog window will open for the assignment of the instru-
ment.

Check solutions
Check whether the solutions required in the instrument-specific com-
mands are present and available.

Check sensors
Check whether the sensors required in the instrument-specific com-
mands are present and available.

Note

When it is in READY status, the start test can also be triggered manually
with the menu item Tools

Run test.

4 - Conditioning (only for methods with conditioning)
Conditioning is started in the case of methods which contain commands
with switched-on conditioning. The requirement for this is that the Auto-
matic conditioning
option is switched on in the START command.
Conditioning is not yet part of determination and causes no counters to be
incremented (Sample number, Statistics).

Conditioning condition not fulfilled
After conditioning starts, the status switches to COND BUSY. The deter-
mination can be started in this status with [Start], even though the
conditioning requirement is not fulfilled. Conditioning can also be can-
celed with [Stop] or interrupted with [Hold] and resumed with [Cont].
If an error occurs during conditioning, then neither the error track nor
the exit track will be run through.

Advertising