5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

1083

Lower limit pH

Lower limit of the measured value. When the measured value falls below
this value, the event Measured value lower limit exceeded is triggered.

Titrando, 855
Range

-20.000 ... 20.000

Default value

-20.000

Titrino
Range

-20.00 ... 20.00

Default value

-20.00

Lower hysteresis pH

Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value, then the event Measured value lower
limit ok
is triggered.

Titrando, 855
Range

0 ... 20.000

Default value

20

Upper limit pH

Upper limit of the measured value. If the measured value exceeds this limit,
then the event Measured value upper limit exceeded is triggered.

Titrando, 855
Range

-20.000 ... 20.000

Default value

20.000

Titrino
Range

-20.00 ... 20.00

Default value

20.00

Upper hysteresis pH

The upper hysteresis of the measured value. If the measured value once
again falls below the upper limit by this hysteresis value, then the event
Measured value upper limit ok will be triggered.

Titrando, 855
Range

0 ... 20.000

Default value

20

Action

Selection of the event to be triggered if the upper or lower limit is exceeded: