5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

466

■■■■■■■■

tiamo 2.2 (ProcessLab)

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL

5.6.3.2.3

DET U

5.6.3.2.3.1

DET U - Overview

Dialog window: Method

▶ DET U ▶ Properties... ▶ DET U - 'Command name'

Command for dynamic equivalence point titrations with potentiomet-
ric voltage measurement.

Devices
This command can be executed with the following devices:

Titrando: 808, 809, 835, 836, 857, 888, 904, 905, 906, 907

Titrino: 716, 721, 736, 751, 785, 794, 798, 799

Robotic Titrosampler: 855

Appearance
The command has the following appearance:

Parameters
The parameters for the command DET U are configured in the following 7
tabs:

■

General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.

■

Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titration.

■

Titration parameters
Parameters for the run of the titration.

■

Stop conditions
Definitions of conditions which cause the stop of the titration.

■

Potentiometric evaluation
Parameters for the potentiometric evaluation of titration curves.

■

Additional evaluations
Definition of additional methods for the evaluation of the titration curves.