2 General program functions

tiamo 2.2 (ProcessLab)

■■■■■■■■

35

Identifica-
tion

Description

Commands

.CYL

Cylinder volume of the exchange or
dosing unit used for the command

DET, MET, SET,
KFT, STAT,
STDADD dos,
STDADD auto,
ADD, DOS, LQH,
PREP, EMPTY

.DBL

Total duration for the processing of
the command in s

DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS,
CAL MEAS,
STDADD, DOS

.DEF

Default value which is defined in the
configuration for an IO port

ANALOG OUT,
DIGITAL OUT

.DRI

Current and/or last drift for drift cor-
rection in µL/min

SET, KFT, KFC,
BRC

.DSC

Time for processing all start condi-
tions s

DET, MET, SET,
KFT, STAT

.DTI

Time for drift correction (time from
the start of the titration to the end
of the command) in s

SET, KFT, KFC,
BRC

.EGF

Last measured gas flow rate (meas-
ured value following processing of
the command) in mL/min

MEAS T/FLow

.EME

End measured value (measured
value after processing of the com-
mand) in the unit of the measured
value

DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS
(without REF,
Spec), CAL
MEAS, STDADD,
DOS

.ENP

Electrode zero point for the sensor
used for the command (dimension-
less for pH sensor or in mV for the
ISE sensor) and/or electrode zero
point calculated from the calibra-
tion (for STDADD and CAL LOOP)

DET pH, DET U,
MET pH, MET U,
SET pH, SET U,
STAT, MEAS pH,
MEAS U, MEAS
T, MEAS Conc,
STDADD, CAL
LOOP, DOS

.EP{x}.CHA

Load for endpoint x (1…9) in mA·s

KFC, BRC