5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

734

■■■■■■■■

tiamo 2.2 (ProcessLab)

[Properties]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 880)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.3.4.6.10

SET Upol - Fix endpoint evaluation

Dialog window: Method

▶ SET Upol ▶ Properties... ▶ Additional evaluations ▶ Fix

endpoint evaluation

▶ [New]/[Properties] ▶ Fix endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-200.0 ... 200.0 µA

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL

5.6.3.5

KFT

5.6.3.5.1

KFT - Overview

Menu item: Method

▶ Paste ▶ New command...

Command for volumetric Karl Fischer Titrations.

Principle
With this command, volumetric titrations for determining the water content
by the Karl Fischer method with automatic conditioning are carried out. A
typical feature of the Karl Fischer titration is the controlled reagent addition,
which depends on the control difference (current measured value – end-
point), with the aim of reaching the measured value given as being the
predefined target endpoint as quickly and accurately as possible. Overtitrat-
ing is avoided to a great extent. The titration termination at the endpoint
takes place either drift-controlled or after a waiting period. The volume
added up to the endpoint gives the calculable reagent consumption.