7 set upol - conditioning, Conditioning – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 745

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

729

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

5.6.3.4.6.7

SET Upol - Conditioning

Tab: Method

SET Upol Properties... Conditioning

Command name

Name of the command.

Input

25 characters

Conditioning

on | off (Default value: off)

Switching conditioning on/off. If conditioning is deactivated all following
parameters are not visible.

Start drift

The determination can only be started when the currently measured drift is
less than the entered start drift entered here. If this is the case, then the
status displayed in the Run subwindow in the Workplace program part
changes to COND READY.

Titrando, 855
Range

1 ... 999 µL/min

Default value

20 µL/min

Drift correction

If this parameter is set to automatic or manual the drift correction ( = Drift
value
× Time for drift correction) caused by empty consumption during
titration is deducted from the total consumption. The Time for drift cor-
rection
is the period between end of conditioning and end of titration.

Selection

automatic | manual | off

Default value

off

automatic
The drift lastly determined during conditioning will be used as drift value.
manual
The drift value defined under Drift value will be used.