10 det u - volume window – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 506

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

490

■■■■■■■■

tiamo 2.2 (ProcessLab)

Titrando, 855
Range

0 ... 200

Default value

5

EP recognition

Filter for the recognition of equivalence points:

Titrando, 855
Selection

first | greatest | last | ascending | descending

Default value

first

first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be recog-
nized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.

5.6.3.2.3.10

DET U - Volume window

Dialog window: Method

DET U Properties... Potentiometric evalua-

tion

Evaluation with volume window (ml) [New]/[Properties] Volume window

#

Volume windows are defined sub-regions (windows) on the volume axis for
which different parameters for the potentiometric evaluation can be
defined. Only those endpoints which lie within these windows and addi-
tionally meet the parameters defined for each window will be recognized.
Only one endpoint will be recognized per window.

Lower limit

Lower limit for the volume window.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

0.00000 mL

Upper limit

Upper limit for the volume window.