5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

615

Evaluation with volume window (mL)

[New]

Opens the dialog window Volume window # in which the parameters for
a new window can be entered (see Chapter 5.6.3.3.4.10, page 621).

[Properties]

Opens the dialog window Volume window # in which the parameters for
the selected window can be edited (see Chapter 5.6.3.3.4.10, page
621)
.

[Delete]

Deletes the window selected in the table.

5.6.3.3.4.7

MET Ipol - Additional evaluations

Tab: Method

▶ MET Ipol ▶ Properties... ▶ Additional evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for evaluation of titration curves can be
activated and defined on this tab:

Fix endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this option is enabled, then the associated values for the other measured
quantities will be interpolated with a fixed measured quantity (measured
value
, time or volume) for the fix endpoint from the measuring point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Opens the dialog window Fix endpoint evaluation # to enter a new fix
endpoint (see Chapter 5.6.3.3.4.11, page 622).

[Properties]

Opens the dialog window Fix endpoint evaluation # to edit the fix end-
point selected in the table (see Chapter 5.6.3.3.4.11, page 622).