5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

581

If the start measured value is achieved by the addition of a start volume,
then titration will start directly.

Range

-2000.0 ... 2000.0 mV

Selection

off

Default value

off

Dosing rate

Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.

Range

0.01 ... 166.00 mL/min

Default value

5.00 mL/min

Selection

maximum

Start slope

Note

Is displayed only for Titrando and 855.

Start slope

When the start slope is reached the dosing of the start volume is stopped
and the titration is started. If the start slope is achieved by the dosing of a
start volume then the titration starts directly.

Range

0 ... 999 mV/mL

Selection

off

Default value

off

Dosing rate

Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.

Range

0.01 ... 166.00 mL/min

Default value

5.00 mL/min

Selection

maximum

Pause

Pause

Waiting time, e.g. for the electrode to settle down after the start or a reac-
tion time after the addition of the start volume. The pause follows at the
end of all the start conditions.