5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

446

■■■■■■■■

tiamo 2.2 (ProcessLab)

Sensor

Selection of a sensor of type pH electrode from the list of sensors available
in the sensor table. The calibration data for the sensor will be adopted for
the determination.

Selection

Sensor name | pH electrode

Default value

pH electrode

Temperature measurement

Type of temperature measurement.

Titrando, 855
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the temperature entered manually under
Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.

Stirrer

Stirrer

Selection of the stirrer.

Titrando, 855
Selection

1 | 2 | 3 | 4 | off

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Titrando, 855
Range

-15 ... 15

Default value

8