5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

459

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and temperature
for the minimum measured value will be interpolated from the measuring
point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.1 ... 20.0 pH/mL

Default value

1.0 pH/mL

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and temperature
for the maximum measured value will be interpolated from the measuring
point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.1 ... 20.0 pH/mL

Default value

1.0 pH/mL

Break-point evaluation

Note

This evaluation method is possible only with Titrando and 855.

Break-point evaluation

on | off (Default value: off)

A break-point evaluation is used to determine sharp changes of direction in
the titration curve.

EP criterion

Measure of the minimum sharpness of the break-point. The smaller the EP
criterion set, the more break-points will be found. As this is a relative value