5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

936

■■■■■■■■

tiamo 2.2 (ProcessLab)

Parameters
The parameters for the command MEAS T are configured on the following
4 tabs:

■

General/Hardware
Parameters for devices, sensors and stirrers.

■

Measuring parameters
Parameters for setting the measurement procedure.

■

Evaluations
Defining further methods for evaluating measurement curves.

■

Additional measured values
Defining further measured values which can be saved as additional col-
umns in the measuring point list.

Command variables
The following command variables are generated in the method run by the
command MEAS T and can be used in formulas under the designation
'Command name.Variable designation':

Identification

Description

.BP{x}.MEA

Measured value (temperature) for the break point ×
(1…9) in the unit of the measured value

.BP{x}.TEM

Temperature for the break point × (1…9) in °C

.BP{x}.TIM

Time for the break point × (1…9) in s

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.DBL

Total duration for the processing of the command in s

.EME

End measured value (measured value after processing
of the command) in the unit of the measured value

.ENP

Electrode zero point of the sensor used for the com-
mand (dimensionless)

.ETE

End temperature (temperature following processing of
the command) in °C

.FIN

Command status; 1 = Command has ended at least
once
, 0, invalid (variable not available) = Command
has never ended

.FP{x}.MEA

Measured value (temperature) for the fixed endpoint ×
(1…9) in the unit of the measured value

.FP{x}.TEM

Temperature for the fixed endpoint × (1…9) in °C