BPS / MX / RS Series SCPI Programming Manual
100
Manual P/N 7003-961 Rev AA
4.20
Source Subsystem - PONSetup
This subsystem will control the initial condition of the power source at the power on state.
Subsystem syntax
[SOURce:]
PONSetup:
ALControl 0|1|OFF|ON
Set initial ALC state.
CLOCk INT|EXT
Set the the clock source to internal or external
CURRent <n>
Set the current limit
FREQuency <n>
Set the frequency value
NOUTput
ONEPhase|THReephase|LAST
Set
the
command
language
OLOad
[:MODE]
CVOL|CCUR Set
the
over load mode const voltage or const
current
OUTPut
[:RELay]
0|1|OFF|ON
Set
the state of the output relay
PHASe
[:ANGLe]
Set the phase angle relative to external
SENSe INT|EXT
Set the voltage sense to internal or external
VOLTage
[:LEVel] <n>
Set the voltage level
MODE DC|AC|ACDC
Set the voltage mode
VRANge 150|300|400
Set the voltage range
WGRoup <n>
Set the wave group to 0...3
[3Pi Controller
Only]
PONSetup:ALControl
This command sets the initial ALC mode to either On or Off at power on.
Command Syntax
[SOURce:]PONSetup:ALControl <NRf+>
Parameters
0 | 1 | OFF | ON
Examples PONS:ALC
0
Query Syntax
PONS:ALC?
Returned Parameters
0 | 1
PONSetup:CLOCk
This command determines the source of its clock at the power on. if internal, the source
uses its internal clock. if external, a clock source must be supplied on the appropriate input.
Command Syntax
[SOURce:]PONSetup:CLOCk <source>
Parameters INTernal|EXTernal
Examples PONSetup:CLOCk
INT
Query Syntax
PONSetup:CLOCk?
Returned Parameters
<CRD>
Related Commands
FREQ:MODE