AMETEK Programmable Power

BPS / MX / RS Series SCPI Programming Manual

100

Manual P/N 7003-961 Rev AA

4.20

Source Subsystem - PONSetup

This subsystem will control the initial condition of the power source at the power on state.

Subsystem syntax

[SOURce:]
PONSetup:

ALControl 0|1|OFF|ON

Set initial ALC state.

CLOCk INT|EXT

Set the the clock source to internal or external

CURRent <n>

Set the current limit

FREQuency <n>

Set the frequency value

NOUTput

ONEPhase|THReephase|LAST

Set

the

command

language

OLOad
[:MODE]

CVOL|CCUR Set

the

over load mode const voltage or const

current

OUTPut

[:RELay]

0|1|OFF|ON

Set

the state of the output relay

PHASe

[:ANGLe]

Set the phase angle relative to external

SENSe INT|EXT

Set the voltage sense to internal or external

VOLTage

[:LEVel] <n>

Set the voltage level

MODE DC|AC|ACDC

Set the voltage mode

VRANge 150|300|400

Set the voltage range

WGRoup <n>

Set the wave group to 0...3

[3Pi Controller

Only]

PONSetup:ALControl

This command sets the initial ALC mode to either On or Off at power on.

Command Syntax

[SOURce:]PONSetup:ALControl <NRf+>

Parameters

0 | 1 | OFF | ON

Examples PONS:ALC

0

Query Syntax

PONS:ALC?

Returned Parameters

0 | 1

PONSetup:CLOCk

This command determines the source of its clock at the power on. if internal, the source
uses its internal clock. if external, a clock source must be supplied on the appropriate input.

Command Syntax

[SOURce:]PONSetup:CLOCk <source>

Parameters INTernal|EXTernal
Examples PONSetup:CLOCk

INT

Query Syntax

PONSetup:CLOCk?

Returned Parameters

<CRD>

Related Commands

FREQ:MODE