AMETEK BPS Series Programming Manual User Manual
Page 125

BPS / MX / RS Series SCPI Programming Manual
AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA
125
TRACe:CATalog
These commands return a list of defined waveform names. The list includes both pre-
defined waveforms such as SINusoid, SQUare, and CSINusoid, as well as any user-defined
waveforms.
Query Syntax:
TRACe:CATalog?
Returned Parameters
<SRD>
Example TRAC:CAT?
Related Commands
TRAC:DATA TRAC:DEL FUNC:SHAP
TRACe:DEFine
These commands define a new waveform with the name <waveform_name> and allocates
storage for its data. The waveform name can then be referenced by the TRACe:DATA.
Note: The TRAC:DEF command causes waveform catalog data to be writing to the
Flash memory of the unit. This process requires about 500 msec to complete.
The TRACE:DATA command which normally follows should not be sent
during this period to allow this process to complete or the first part of the data
may be missed and an error will result.
Command Syntax:
TRACe:DEFine <waveform_name>
Parameters <waveform_name>
Example TRAC:DEF
flattop
Related Commands
TRAC:DATA TRAC:DEL FUNC:SHAP
TRACe:DELete
This command deletes the user-defined waveform table with the name <waveform_name>
and makes its memory available for other waveforms.
Command Syntax:
TRACe:DELete[:NAME]<waveform_name>
Parameters <waveform
name>
Example TRAC:DEL
flattop
Related Commands
TRAC:DATA TRAC:DEL FUNC:SHAP
TRACe:DELete
This command deletes ALL user-defined waveforms at once and makes all waveform
memory available for other waveforms. This command is only supported by firmware
revisions 0.16 and higher.
Command Syntax:
TRACe:DELete:ALL
Parameters <waveform
name>
Example TRAC:DEL:ALL
Related Commands
TRAC:DATA TRAC:DEL FUNC:SHAP