AMETEK Programmable Power

BPS / MX / RS Series SCPI Programming Manual

178

Manual P/N 7003-961 Rev AA

:PAUSe

Set harmonics pause.

:TABle

Select Individual harmonic table 1,2,3 or ALL.

:IHARmonics

:LEVel

Set

interharmonics

level.

:DWELl

Set

interharmonics

dwell.

:PAUSe

Set interharmonics pause.

IEC413:STATe

This command will enable the IEC413 and initialize the associated data. Note that user
defined test levels for the USER group can only be programmed when the IEC413 state is
OFF.

Command Syntax

IEC413:STATe <state>

Parameters

<ON|OFF|1|0 >

Examples

IEC413:RUN STAT ON

Query Syntax

IEC413:STAT?

Returned Parameters

<boolean>

IEC413:RUN

This command will run the test. The parameters will define the section of test to be run.

Command Syntax

IEC413:RUN <group>

Parameters

< ALL|GROup|RANGe|POINt >

Examples IEC413:RUN

ALL

Query Syntax

IEC413:RUN?

Returned Parameters

<group>

IEC413:PAUSe

This command will cause the test to pause at the time the command is issued.

Command Syntax

IEC413:PAUSe <boolean>

Parameters ON|OFF|1|0
Examples IEC413:PAUS

ON

Query Syntax

IEC413:PAUS?

Returned Parameters

<boolean>

IEC413:CLASs

This command will select the class. There are two predefined classes 2 and 3. Class 1 is
supported through the use of the USER defined class. Other tests levels can be set in the
USER class as well.

Command Syntax

IEC413:CLASs <class>

Parameters 2|3|USER
Examples IEC413:CLAS

3

Query Syntax

IEC413:CLAS?

Returned Parameters

<class>