BPS / MX / RS Series SCPI Programming Manual

AMETEK Programmable Power

Manual P/N 7003-961 Rev. AA

213

Appendix C: Error Messages

Any errors that occur during operation from either the front panel or the remote control
interface will result in error messages. Error messages are displayed in the upper left hand
corner of the LCD display. They are also stored in the error message queue from which they
can be queried using the SYST:ERR? Query. The error queue has a finite depth. If more
error messages are generated than can be held in the queue, a queue overflow message
will be put in the last queue location. To empty the queue, use the error query until the No
Error result is received.

Errors appearing on the LCD will generally remain visible until the user moves to another
screen. If multiple error messages are generated in succession, only the last message will
be visible as there is only space for one error message on the LCD display.

The same area of the display is also used to display status messages. While error
messages always have a negative error number, status messages have a positive number.

The table below displays a list of possible error and status messages along with their
possible cause and remedy.

Number

Message String

Cause

Remedy

0

"No error"

No errors in queue

-100

"Command error"

Unable to complete
requested operation

Unit may be in a mode inconsistent
with request.

-102

"Syntax error"

Command syntax
incorrect.

Misspelled or unsupported command

-103 "Invalid

separator"

SCPI separator not
recognized

See SCPI section of programming
manual.

-104

"Data type error"

Data type invaled.

Check command for supported data
types

-108 "Parameter

not

allowed"

One or more
additional parameters
were received.

Check programming manual for
correct number of parameters

-109 "Missing

parameter"

Too few parameters
received for
requested operation

Check programming manual for
correct number of parameters

-110 "Command

header

error"

Command header
incorrect

Check syntax of command.

-111 "header

separator

error"

Invalid command
separator used.

Use semi-colon to separate
command headers

-112 "Program

mnemonic

too long"

Syntax error

Check programming manual for
correct command syntax

-113

"Undefined header"

Command not
recognized error

Check programming manual for
correct command syntax

-120

"Numeric data error"

Data received is not a
number

Check programming manual for
correct command syntax

-121

"Invalid character in
number"

Number received
contains non-numeric
character(s)

Check programming manual for
correct command syntax

-123

"Exponent too large"

Exponent in number
exceeds limits

Check programming manual for
correct parameter range

-128

"Numeric data not
allowed"

Number received
when number is not
allowed.

Check programming manual for
correct command syntax