FREQuency:SLEW:TRIGgered
This command sets the rate at which frequency changes during a triggered output transient.
Instantaneous frequency changes can be obtained by sending MAXimum.
Command Syntax
[SOURce:]FREQuency:SLEW:TRIGgered <NRf+>
Parameters
0.01 to 1E9
⏐MAXimum
Unit
Hz (Hertz per second)
*RST Value
MAXimum
Examples
FREQ:SLEW:TRIG 75 FREQ:SLEW:TRIG MAX
Query Syntax
[SOURce:]FREQuency:SLEW:TRIG?
Returned Parameters
<NRf>
Related Commands
FREQ:SLEW:MODE
FREQ
FREQuency:TRIGgered
This command programs the frequency that the output will be set to during a triggered step
or pulse transient.
Command Syntax
[SOURce:]FREQuency:TRIGgered <NRf+>
Parameters
0 to maximum frequency range specified by the LIMit:FREQuency
command
Unit Hz
(Hertz)
*RST Value
60 Hz
Example FREQ:TRIG
50
Query Syntax
[SOURce:]FREQuency:TRIGgered?
Returned Parameters
<NR2>
Related Commands
FREQ FREQ:MODE