BPS / MX / RS Series SCPI Programming Manual

AMETEK Programmable Power

Manual P/N 7003-961 Rev. AA

43

CALibrate:MEASure:CURRent:TEMPerature:DC[:FSCale] <NRf>

This command initiates the calibration of the DC current measurement at elevated
temperature.

Command Syntax

CALibrate:MEASure:CURRent:TEMPerature:DC[:FSCale]

Parameters

<NRf> (0 or desired offset value)

Examples CAL:MEAS:CURR:TEMP:DC
Query Syntax

CALibrate:MEASure:CURRent:TEMPerature:DC?

Returned Parameters

<NR1> (value range -1000 to + 1000

Related Commands

CALibrate:MEASure:CURRent[:AMB]:DC[:FSCale]

CALibrate:MEASure:CURRent:TEMPerature:DC:ZERO <NRf>

This command initiates the offset adjustment of the DC current measurement at elevated
temperature.

Command Syntax

CALibrate:MEASure:CURRent:TEMPerature:DC:ZERO

Parameters

<NRf> (0 or desired offset value)

Examples CAL:MEAS:CURR:TEMP:DC:ZERO
Query Syntax

CALibrate:MEASure:CURRent:TEMPerature:DC:ZERO?

Returned Parameters

<NR1> (value range 0 to +5)

Related Commands

CALibrate:MEASure:CURRent[:AMB]:DC:ZERO