BPS / MX / RS Series SCPI Programming Manual
AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA
49
4.2.8
Output - DC Voltage
CALibrate[:SOURce]:VOLTage:DC:LRANge[:FSCale]<NRf+>
This command will result in the calibration coefficent being calculated for the DC full-scale
output voltage in the low voltage range.
Command Syntax
CALibrate:VOLTage:DC:LRANge <NRf+>
Parameters
<NRf> (programmed output voltage)
Examples CAL:VOLT:DC:LRAN
-2
Query Syntax
CALibrate:VOLTage:DC:LRANge?
Returned Parameters
<NRf> (cal coefficient value between 0.9 and 1.1)
Related Commands
CAL:VOLT:DC:LRAN:NEG
CALibrate[:SOURce]:VOLTage:DC:LRANge:ZERO <NRf+>
This command will set the calibration coefficent for the dc output voltage offset at the low
voltage range.
Command Syntax
CALibrate:VOLTage:DC:LRANge:ZERO <NRf+>
Parameters
<NRf> (a value between -127 and +128)
Examples CAL:VOLT:DC:LRAN:ZERO
+10
Query Syntax
CALibrate:VOLTage:DC:LRANge:ZERO?
Returned Parameters
<NR1> (value range -127 to +128)
Related Commands
CAL:VOLT:DC:LRAN:ZERO
CALibrate[:SOURce]:VOLTage:DC:HRANge[:FSCale]<NRf+>
This command will result in the calibration coefficent being calculated for the DC full scale
output voltage in the high voltage range.
Command Syntax
CALibrate:VOLTage:DC:HRANge <NRf+>
Parameters
<NRf> (programmed output voltage)
Examples CAL:VOLT:DC:HRAN
-2
Query Syntax
CALibrate:VOLTage:DCHRANge?
Returned Parameters
<NRf> (cal coefficient value between 0.9 and 1.1)
Related Commands
CAL:VOLT:DC:LRAN:ZERO
CALibrate[:SOURce]:VOLTage:DC:HRANge:ZERO <NRf+>
This command will set the calibration coefficent for the dc output voltage offset at the high
voltage range.
Command Syntax
CALibrate:VOLTage:DC:HRANge:ZERO <NRf+>
Parameters
<NRf> (a value between -127 and +128)
Examples CAL:VOLT:DC:HRAN:ZERO +10
Query Syntax
CALibrate:VOLTage:DC:HRANge:ZERO?
Returned Parameters
<NR1> (value range -127 to +128)
Related Commands
CAL:VOLT:DC:LRAN