AMETEK BPS Series Programming Manual User Manual
Page 105

BPS / MX / RS Series SCPI Programming Manual
AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA
105
Parameter Set
Action
DCYCle
PERiod
WIDTh
√
Sets WIDTh. If WIDTh < PERiod, recalculates DCYCle;
otherwise, recalculates the PERiod and DCYCle.
√
Sets PERiod. If WIDTh < PERiod, recalculates DCYCle;
otherwise, recalculates the PERiod and DCYCle.
√
Sets DCYCle and recalculates PERiod
Table 4-1 : PULSe:HOLD = WIDTh parameters
Parameter Set
Action
DCYCle
PERiod
WIDTh
√
Sets WIDTh and recalculates the PERiod
√
Sets PERiod and recalculates the WIDTh
√
Sets DCYCle and recalculates PERiod
Table 4-2 : PULSe:HOLD = DCYCle parameters
PULSe:DCYCle
This command sets the duty cycle of the triggered output pulse. The duty cycle units are
specified in percent.
Command Syntax
[SOURce:]PULSe:DCYCle<NRf+>
Parameters
0 to 100%|MINimum|MAXimum
*RST Value
50%
Examples PULS:DCYC
75
PULS:DCYC
MAX
Query Syntax
[SOURce:]PULSe:DCYCle?
Returned Parameters
<NR2>
Related Commands
PULS:COUN
PULS:PER
PULS:WIDT
PULSe:PERiod
This command sets the period of a triggered output transient The command parameters are
model-dependent.
Command Syntax
[SOURce:]PULSe:PERiod<NRf+>
Parameters
2 msecs to 90,000 secs|MINimum|MAXimum
Unit S
(seconds)
*RST Value
1 sec
Examples
PER 0.001PER MIN
Query Syntax
[SOURce:]PERiod?
Returned Parameters
<NR2>
Related Commands
PULS:COUN
PULS:DCYC
PULS:WIDT