AMETEK BPS Series Programming Manual User Manual
Page 82

AMETEK Programmable Power
BPS / MX / RS Series SCPI Programming Manual
82
Manual P/N 7003-961 Rev AA
FREQuency:SLEW
This command sets the rate at which frequency changes for all programmed changes in
output frequency. Instantaneous frequency changes can be obtained by sending
MAXimum.
Command Syntax
[SOURce:]FREQuency:SLEW[:IMMediate] <NRf+>
Parameters
0.01 to 1E9
⏐MAXimum
Unit
Hz (Hertz per second)
*RST Value
MAXimum
Examples
FREQ:SLEW:IMM 75FREQ:SLEW MAX
Query Syntax
[SOURce:]FREQuency:SLEW?
Returned Parameters
<NRf>
Related Commands
FREQ:SLEW:MODE
FREQ
FREQuency:SLEW:MODE
This command determines how the frequency slew rate is controlled during a triggered
output transient. The choices are:
FIXed
The frequency slew rate is unaffected by a triggered output transient.
STEP
The frequency slew rate is programmed to the value set by
FREQuency:TRIGgered when a triggered transient occurs.
PULSe
The frequency slew rate is changed to the value set by
FREQuency:TRIGgered for a duration determined by the pulse commands.
LIST
The frequency slew rate is controlled by the frequency list when a triggered
transient occurs.
Command Syntax
[SOURce:]FREQuency:SLEW:MODE<mode>
Parameters
FIXed | STEP | PULSe | LIST
*RST Value
FIXed
Examples FREQ:SLEW:MODE
FIX
Query Syntax
[SOURce:]FREQuency:SLEW:MODE?
Returned Parameters
<CRD>
Related Commands
FREQ FREQ:SLEW:TRIG