AMETEK BPS Series Programming Manual User Manual
Page 108

AMETEK Programmable Power
BPS / MX / RS Series SCPI Programming Manual
108
Manual P/N 7003-961 Rev AA
VOLTage
Phase Selectable
This command programs the AC rms, DC or offset output voltage level of the power source.
The voltage mode must be set for AC to accept an AC voltage. The Voltage mode must also
be set to AC+DC to accept a DC offset value.
Command Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude][:AC] <NRf+>
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]:DC] <NRf+>
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]:OFFSet<NRf
+>
Parameters
0 to maximum voltage range specified by the LIMit:VOLTage
command
Unit
V (rms voltage)
*RST Value
0 volt
Examples
VOLT 250
VOLT:LEV 25
Query Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Returned Parameters
<NR2>
Related Commands
[SOUR:]MODE VOLT:TRIG
VOLTage:TRIGgered
Phase Selectable
This command selects the AC rms or DC amplitude that the output voltage will be set to
during a triggered step or pulse transient.
Command Syntax
SOURce:]VOLTage[:LEVel]:TRIGgered:AMPLitude]<NRf+>
Parameters
0 to maximum voltage range specified by the LIMit:VOLTage
command
Unit
V (rms voltage)
*RST Value
0 volt
Examples
VOLT:TRIG 120
VOLT:LEV:TRIG 120
Query Syntax
SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
Returned Parameters
<NR2>
Related Commands
VOLT VOLT:MODE