CMG-6TD

Installing the 6TD

exposed on the instrument's main connector. If you have problems
connecting to the 6TD over a network, you can attach a diagnostics cable,
CAS-PEP-0041, to this port and use Scream! to access the console.
Channel 2 (port 10,002) is connected to the 6TD's digital output, unless
you have connected a serial data cable from the breakout box to a
computer. If the breakout box is connected, the 6TD will send data
streams through that interface rather than to the Wi-Port-NR.
Click on Channel 2 – Serial Settings.

Set the Baud Rate to 19200. This is the default baud rate for the 6TD's

digital output.

Note: If you change the baud rate in Scream! or using the

terminal, you must come back to this page and change the
Baud Rate setting.

The remaining settings can be left at their default values. Click

to save your changes.

For full information on the Wi-Port-NR's configuration options, please
refer to the Wi-Port documentation, which is available on the Lantronix

Web site,

http://www.lantronix.com/

.

3. When you have finished setting up the Wi-Port-NR, apply the new

settings by clicking

. The Wi-Port-NR will

re-boot with the new settings in effect.

Note: Simply changing settings in the web page and clicking

does not effect any functional change. You must

click

before the changes take effect.

If the Wi-Port-NR is using an automatically chosen random IP

(beginning with 169.254), the IP address will change when you do this.
You will need to go back to DeviceInstaller to find out the new IP

address.

MAN-T60-0002

36

Issue F - February 2014

Advertising