Calibrating the 6TD

3. Open a Waveview window on the calibration signal and the returned

streams by holding

while selecting each of them and then

double-clicking. The streams should display the calibration signal

combined with the sensors' own measurements. If you cannot see the
calibration signal, zoom into the Waveview using the scaling icons at
the top left of the window ( )or the cursor keys

and

.

If necessary, drag the calibration stream Cn up the Waveview window,
so that it is at the top.

4. If the returning signal is saturated, retry using a calibration signal with

lower amplitude, until the entire curve is visible in the Waveview

window.

5. If you need to scale one, but not another, of the traces, right-click on

the trace and select Scale…. You can then type in a suitable scale

factor for that trace.

6. Pause the Waveview window by clicking on the

icon.

7. Hold down

and drag across the window to select the calibration

signal and the returning component(s). Release the mouse button,
keeping

held down. A menu will pop up. Choose Broadband

Noise Calibration.