Guralp Systems CMG-6TD User Manual

Page 59

Advertising
background image

CMG-6TD

Configuration with Scream!

diagnostic information, known as multiplexed (“Mux”) channels. In the 6TD,
three channels are used to report the sensor mass position, one measures the

internal temperature of the digitiser and one is used for the returning
calibration signal (see Chapter 6 on page 72).

The collection and transmission of Mux channels is controlled using the Mux
Channels pane:

If a tick is placed in the check-box next to a channel, its data will be collected
and transmitted as a data stream in GCF format, just as with the normal data

channels. To indicate that the data come from a Mux channel, the Stream ID
will take the form ****Mx, where M stands for Mux and x is a hexadecimal

integer (i.e. 0 – 9, and A – F for 10 through 15). The Z, N/S and E/W Mass
Position Mux channels appear as M8, M9 and MA respectively. ME is used for

the temperature sensor and MB for the returning calibration signal.

5.1.5 Ports

The Baud Rates pane of the Configuration window allows you to program the

baud rate and stop bits for the 6TD's output port.

Caution: If you have a 6TD with Ethernet or Wi-Fi options, the

settings you configure here are used both on the standard data
output port, and on the internal port which sends data to the

Ethernet/Wi-Fi module. If you change them, you will also need to
configure the Ethernet/Wi-Fi module to receive data with the new

settings. This can be done using the Lantronix DeviceInstaller
utility (see section 4.10 on page 32 (wired) or section 4.11 on page

37 (wireless).

MAN-T60-0002

59

Issue F - February 2014

Advertising