6calibrating the 6td, 1 the calibration pack, 2 poles and zeroes – Guralp Systems CMG-6TD User Manual

Page 72

Advertising
background image

CMG-6TD

Calibrating the 6TD

6

Calibrating the 6TD

6.1

The calibration pack

All Güralp sensors are fully calibrated before they leave the factory. Both
absolute and relative calibration calculations are carried out. The results are

given in the calibration pack supplied with each instrument:

Works Order : The Güralp factory order number including the
instrument, used internally to file details of the sensor's manufacture.

Serial Number : The serial number of the instrument

Date : The date the instrument was tested at the factory.

Tested By : The name of the testing engineer.

There follows a table showing important calibration information for each
component of the instrument, VERTICAL, NORTH/SOUTH, and EAST/WEST.

Each row details:

Velocity Output (Differential) : The sensitivity of each component to
velocity at 1 Hertz, in volts per ms

-1

. Because the 6TD uses differential

outputs, the signal strength as measured between the +ve and –ve lines
will be twice the true sensitivity of the instrument. To remind you of

this, the sensitivities are given as 2 × (single-ended sensitivity) in each
case.

Mass Position Output : The sensitivity of the mass position outputs to

acceleration, in volts per ms

-2

. These outputs are single-ended and

referenced to signal ground.

Feedback Coil Constant : A constant describing the characteristics of

the feedback system. You will need this constant, given in amperes per
ms

-2

, if you want to perform your own calibration calculations (see

below).

Power Consumption : The average power consumption of the sensor
during testing, given in terms of the current (in amperes) drawn from a

12 Volt supply.

Calibration Resistor : The value of the resistor in the calibration
circuit. You will need this value if you want to perform your own

calibration calculations (see below).

6.2

Poles and zeroes

Most users of seismometers find it convenient to consider the sensor as a
“black box”, which produces an output signal V from a measured input x. So

long as the relationship between V and x is known, the details of the internal

MAN-T60-0002

72

Issue F - February 2014

Advertising