Appendix

97

Table 10-5 lists the sub-parameters available when configuring a scale using INDUST mode. The table includes the

default values of the INDUST sub-parameters and the effective (not configurable) values used by NTEP,

CANADA, OIML and NONE regulatory modes

Table 10-5. REGULA/INDUST Mode Parameters, Comparison with Effective Values of Regulatory Modes

REGULA/INDUST Parameter

REGULA Mode

Parameter

Description

INDUST

NTEP

CANADA

OIML

NONE

SNPSHT

Display or Scale weight source

DISPLAY

DISPLAY

DISPLAY

DISPLAY

SCALE

ZTARE

Remove tare on ZERO

NO

NO

NO

YES

YES

KTARE

Always allow keyed tare

YES

YES

NO

YES

YES

MTARE

Multiple Tare Action

REPLAC

REPLAC

NOTHIN

REPLAC

REMOVE

NTARE

Allow negative tare

NO

NO

NO

NO

YES

CTARE

Allow CLEAR tare to clear tare

YES

YES

YES

NO

YES

PRTMOT

Allow print while in motion

NO

NO

NO

NO

YES

PRTPT

Add PT to keyed tare print

NO

NO

YES

YES

NO

OVRBAS

Zero base for overload calculation

CALIB

CALIB

CALIB

SCALE

CALIB

10.6 Data Formats

10.6.1 Stream Serial Data Format
If stream data transmission is configured for the communication ports (

STRLFT

or

STRIND

), by default the 880

sends data using the Rice Lake Weighing Systems serial data format shown in Figure 10-1:

<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>

ASCII 02

decimal

Polarity:
< space > = Positive
<–> = Negative

W eight: 7 digits, right-justified, dummy
zer o es, d ecimal point w ith no leadin g
zer oes except for leading zer o immediately
pr eceding the decimal point. Leading
zer oes transmitted as spaces.

L = pounds
K = kilogram s
T = tons (T , TN )
O = ounces
G = grams
< space > = NONE

G = Gr oss

N = Net

Status:
< space > = valid
I = Invalid
M = Motion
O = Over/under range

<CR> <LF>

or <CR>

Z = Center zero

Figure 10-1. Stream Serial Data Format

SFMT parameter default – <2><P><W7.><U><M><S><CR><LF>

Note

Format can be changed, see Section 10.7.

The character values can be changed for the stream format tokens, see Table 6-10 on page 70.

If the COM port is set to TYPE = RS485, the port will not stream data, and cannot be used in a local/remote
application. See Section 10.6.3
.

10.6.2 Print Output Serial Data Format
When print format is configured for the demand (print) port in the setup menus, the 880 uses a data string formatted

for a basic ticket printout. The particular print format printed depends on the indicator configuration. See

Section 7.0 for print formatting.

Use the EDP commands, Revolution

®

or the front panel to fully customize the print to work with a wide variety of

printers, and other remote equipment.