Calibration

59

4.1.1

Five-point linearization

Five-point linearization (using the WLIN parameter) provides increased scale accuracy by calibrating the indicator

at up to five additional points between the zero and span calibrations.

Linearization is optional: if choosing not to perform linearization, skip the

WLIN

parameter; if linearization values

have previously been entered, these values are reset to zero during calibration of

WZERO

. To perform linearization,

follow the procedure below.

Note

The linearity points must be less than the WSPAN point.

1. With

WLIN

displayed, press

TARE

to go to the first linearization point (

POINT1

).

2. Press again,

WGT1

is displayed.

3. Press to edit the value:

• Press or to select the digit.
• Press or to increment or decrement the value.

• Press

TARE

when the value is correct (the decimal point will be set in the following step).

• Press or to move the decimal point position.

• Press

TARE

when the value is correct. The indicator will display

CAL1

.

4. Place test weights on the scale and press

TARE

. The indicator will display the previously captured A/D

counts for the linearization point.

5. Press

TARE

again to calibrate. The indicator displays

Calibrating, Please Wait

while calibration is in

progress. When complete,

WGT1

is displayed.

6. Press to

POINT1

, then press to

POINT2

.

7. Repeat for up to five linearization points. To exit the linearization parameters, press to return to

WLIN

.

4.1.2

Rezero

The rezero function is used to remove a calibration offset when an apparatus is used to hang the test weights. If no

other apparatus was used to hang the test weights during calibration, remove the test weights and press to return

to the

CALIBR

menu.

1. With

REZERO

displayed, press

TARE

or to access the rezero function.

2. If an apparatus is used during calibration, remove it and the test weights from the scale. The indicator will

display the AD count from the previous zero (

WZERO

) calibration.

3. With all weight removed, press

TARE

to rezero the scale. This function acquires a new ZERO calibration

value. The indicator displays

Calibrating, Please Wait

while the zero and span calibrations are adjusted.

When complete,

LAST

is displayed.

Note

For more information on LAST or TEMP see Section 4.2 or Section 4.3.

4. Press

MENU

to return to weigh mode.