Calibration menu, Ports com menu – Rice Lake 880 Performance Series Indicator/Controller Technical/Service Manual User Manual

Page 74

Advertising
background image

68

880 Technical/Service Manual

Calibration Menu

Table 6-5. CALIBR EDP Commands

Command

Menu

Description

Choices / Range

SC.WZERO#1

WZERO

Perform zero calibration

n/a

SC.WSPAN#1

WSPAN

Perform span calibration

n/a

SC.LC.CD#1

n/a

Raw count at zero

-2147483646 to 2147483647

SC.LC.CW#1

n/a

Raw count at span

-2147483646 to 2147483647

SC.LC.CZ#1

n/a

-2147483646 to 2147483647

SC.REZERO#1

REZERO

Perform zero caibration

n/a

SC.WLIN.C1#1

n/a

Calibrate linearization point 1

n/a

SC.WLIN.C2#1

n/a

Calibrate linearization point 2

n/a

SC.WLIN.C3#1

n/a

Calibrate linearization point 3

n/a

SC.WLIN.C4#1

n/a

Calibrate linearization point 4

n/a

SC.WLIN.C5#1

n/a

Calibrate linearization point 5

n/a

SC.WLIN.F1#1

CAL 1

Raw count value for linearization point 1

-2147483646 to 2147483647

SC.WLIN.F2#1

CAL 2

Raw count value for linearization point 2

-2147483646 to 2147483647

SC.WLIN.F3#1

CAL 3

Raw count value for linearization point 3

-2147483646 to 2147483647

SC.WLIN.F4#1

CAL 4

Raw count value for linearization point 4

-2147483646 to 2147483647

SC.WLIN.F5#1

CAL 5

Raw count value for linearization point 5

-2147483646 to 2147483647

SC.WLIN.V1#1

WGT 1

Test weight value for linearization point 1

0.0 - 999999.0

SC.WLIN.V2#1

WGT 2

Test weight value for linearization point 2

0.0 - 999999.0

SC.WLIN.V3#1

WGT 3

Test weight value for linearization point 3

0.0 - 999999.0

SC.WLIN.V4#1

WGT 4

Test weight value for linearization point 4

0.0 - 999999.0

SC.WLIN.V5#1

WGT 5

Test weight value for linearization point 5

0.0 - 999999.0

SC.WVAL#1

WVAL

Test weight value

0.00001 - 999999.0

Note

The menu items, CAL1 – CAL5 are used to do the calibration. A value cannot be keyed in.

The SC.WLIN.Fx#1 EDP commands can be used to view and edit the value but they do not perform the
calibration. Use the SC.WLIN.Cx#1 commands to perform the calibration.

Ports COM Menu

Table 6-6. PORTS (COM) EDP Commands

Command

Menu

Description

Choices / Range

EDP.BAUD#1

BAUD

Port baud rate

1200, 2400, 4800, 9600, 19200,
28800, 38400, 57600, 115200

EDP.BITS#1

BITS

Port data bits and parity

8NONE, 7EVEN, 7ODD

EDP.ECHO#1

ECHO

Specifies whether characters received by the
port are echoed back to the sending unit

OFF, ON

EDP.EOLDLY#1

EOLDLY

Port end-of-line delay in 0.1 sec intervals

Range: 0 - 255

EDP.TYPE#1

TYPE

Specifies RS-232 or RS-485 communication

232, 485

EDP.ADDRESS#1

ADDRES

RS-485 address

Range: 0 - 255

EDP.PRNMSG#1

PRNMSG

Print message

OFF, ON

EDP.RESPONSE#1

RESPNS

Response

OFF, ON

EDP.SFMT#1

SFMT

Stream format

Alphanumeric, max Length: 200

EDP.STOPBITS#1

STOP B

Stop Bits

1, 2

EDP.TERMIN#1

TERMIN

Termination character

CR/LF, CR

EDP.TRIGGER#1

TRIGGE

Selects the operation of the port

COMAND, STRLFT, STRIND,
REMOTE