Rice Lake 880 Performance Series Indicator/Controller Technical/Service Manual User Manual

Page 38

Advertising
background image

32

880 Technical/Service Manual

SCALE Menu

Parameter

Choices

Description

GRADS

10000
1-100000

Graduations – specifies the number of full scale graduations if SPLIT=OFF. (For multi-range and
multi-interval scales (SPLIT is not

OFF), the GRADS value is derived from the capacity and

display divisions specified for the range or interval.)
The value entered must be in the range 1–100000 and should be consistent with legal
requirements and environmental limits on system resolution.
To calculate GRADS, use the formula: GRADS = Capacity / Display Divisions.
Display divisions are specified under the FORMAT submenu.

SPLIT

OFF
2RNG
3RNG
2INTVL
3INTVL

Multi-range/Interval – specifies whether the scale is full-range (OFF), multi-range (2RNG, 3RNG),
or multi-interval (2INTVL, 3INTVL). For multi-range and multi-interval scales, see the submenu
shown in Figure 3-9 and parameter descriptions in Table 3-6.

FORMAT

Primary
Format

See Figure 3-9 for menu structures. For standard scales see “If SPLIT = OFF”, for multi-range/
interval scales see “If SPLIT = 2RNG, 3RNG, 2INTVL, or 3INTVL”.

ZTRBAN

0
0.0-100

Zero Track Band – automatically zeroes the scale when within the range specified, as long as the
input is within the ZRANGE and scale is at standstill. Specify the zero tracking band in ± display
divisions. Maximum legal value varies depending on local regulations.

ZRANGE

1.900000
0.0-100

Zero Range – selects the range within which the scale can be zeroed. The 1.900000 default
value is ±1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be at
standstill to zero the scale. Maximum legal value varies depending on local regulations.

MOTBAN

1
0-100

Motion Band – sets the level, in display divisions, at which scale motion is detected. If motion is
not detected for the time defined by the standstill parameter, the standstill symbol lights. Some
operations, including print, tare, and zero, require the scale to be at standstill. Maximum legal
value varies depending on local regulations.
If this parameter is set to 0 the standstill annunciator is always lit; operations normally requiring
standstill (zero, tare, print) are performed regardless of scale motion. If 0 is selected, ZTRKBND
must also be set to 0.

OVRLOA

FS+2%
FS+1D
FS+9D
FS

Overload – Determines the point at which the display blanks and an out-of-range error message
is displayed. Maximum legal value varies depending on local regulations.

SSTIME

10
1-65535

Standstill Time – specifies the length of time the scale must be out of motion, in 0.1-second
intervals, before the scale is considered to be at standstill.

DSPRAT

1
1-80

Display update rate. Specifies the display update rate, in the number of 100-millisecond intervals
between updates.

SMPRAT

30HZ
60HZ
120HZ
240HZ
480HZ
960HZ
7.5HZ
15HZ

Sample rate. Selects measurement rate, in samples per second, of the analog-to-digital
converter. Lower sample rate values provide greater signal noise immunity.
Settings of 120Hz or above may be too fast to provide the desired stability in some static
weighing applications.

DFSENS

LIGHT
MEDIUM
HEAVY

Digital filtering sensitivity – the amount of influence the current A/D cycle has on the running
averaged value. The Light setting will respond quicker to an applied weight to immediately
impact the displayed value. Medium and Heavy settings are for applications where weighing
times are longer and expected weight changes are larger.

DFTHRH

0
0-99999

Digital filter cutout threshold – controls the response of the filter and must be set above the noise
disturbances in the system. Value is in grads.

Table 3-5. Scale Menu Parameters