3.0

Configuration

To configure the 880 indicator, the indicator must be placed in setup mode. The setup switch is accessed through a

small hole on the enclosure backplate, insert a non-conductive tool into the access hole and press the configuration

switch. Indicator display changes to show

SCALE

.

Note

Important

Use caution when inserting the non-conductive tool into the backplate, press the tool in about 3/4 inch,
using the board as a guide, until the switch is engaged. Do not use excessive force that may damage the
switch.

If the audit trail is enabled, setup mode may be accessed by pressing

MENU

.

Press

or

until SETUP is

displayed, then press

to SCALE. See Section 2.9.

.

J1

J2

J3

J4 J5 J6

Setup Switch Access

Figure 3-1. Back View – Setup Switch Access

When the indicator is placed in setup mode, the word

SCALE

is shown on the display. The SCALE menu is the first

of eight top-level menus used for configuring the indicator. Detailed descriptions of these menus are given in

Section 3.2.

When configuration is complete, press

MENU

to return to the weigh mode.

3.1

Configuration Methods

The 880 indicator can be configured by using the front panel keys to navigate through a series of configuration

menus or by sending commands or configuration data to any communication port. Configuration using the menus

is described in Section 3.1.3.
Configuration using the communication port can be accomplished using either the EDP command set described in

Section 6.0 or Version 3.0 or later of the Revolution configuration utility.

3.1.1

Revolution Configuration

The Revolution configuration utility provides the preferred method for configuring the 880 indicator. Revolution

runs on a personal computer to set configuration parameters for the indicator (see Section 5.0). When Revolution

configuration is complete, configuration data is downloaded to the indicator.

3.1.2

EDP Command Configuration

The EDP command set can be used to configure the 880 indicator using either a personal computer or terminal.

Like Revolution, EDP command configuration sends commands to the indicator remote EDP port; unlike

Revolution, EDP commands can be sent using any external device capable of sending ASCII characters over a

serial connection.
EDP commands duplicate the functions available using the indicator front panel and provide some functions not

otherwise available. EDP commands can be used to simulate pressing front panel keys, to configure the indicator,

or to dump lists of parameter settings. See Section 6.0 for more information about using the EDP command set.