58

880 Technical/Service Manual

4.1

Front Panel Calibration

1. Put the indicator in setup mode using the setup switch on the back of the unit, see Figure 3-1, (or use

MENU

if audit trail is enabled), and navigate to CALIBR (see Figure 4-1).

2. Press

TARE

or to go to the

WZERO

parameter.

Note

Note

Note

WZERO is used for most normal calibrations with an empty scale. If a special situation exists, where a LAST or
TEMP zero may be used, refer to Section 4.2 or Section 4.3
for more information before performing a WZERO
calibration.

3. Press

TARE

or to view the previously captured A/D count value for zero.

4.

R

emove all weight from the scale platform. If the test weights require hooks or chains, place the hooks or

chains on the scale for zero calibration.

5. Press

TARE

to Calibrate WZERO

.

If calibration of zero is not required, press

MENU

to exit.

6. The indicator displays

CALIBRATING, PLEASE WAIT

while calibration is in progress. When complete,

WVAL

is displayed.

To view the new zero A/D count, repeat Step 3, but instead of pressing Enter while viewing the value, press
Menu to exit.

7. With

WVAL

displayed, press

TARE

to display the stored calibration weight value.

8. To edit the value:

• Press or to select the digit.
• Press or to increment or decrement the value.

• press

TARE

when the value is correct.

• Press or to move the decimal point position.

9. Press

TARE

to store the

WVAL

value and advance to

WSPAN

.

10. With

WSPAN

displayed, press

TARE

or to view the previously captured A/D count value for span.

11. Place test weights on the scale equal to WVAL.

12. Press

TARE

to calibrate

WSPAN

.

Note

Note

If calibration of span is not required, press

MENU

to exit.

13. After

TARE

is pressed, the indicator displays

CALIBRATING, PLEASE WAIT

. When complete

WLIN

is

displayed.

To view the new span A/D count, repeat Step 9, but instead of pressing Enter while viewing the value, press
Menu to exit.

14. When calibration is complete press

MENU

to return to the weigh mode.

Advertising