0 calibration, Calibration – Rice Lake 880 Performance Series Indicator/Controller Technical/Service Manual User Manual

Page 63

Advertising
background image

Calibration

57

4.0

Calibration

The 880 can be calibrated using the front panel, EDP commands, or Revolution.
Calibration consists of the following steps:

• Zero calibration
• Entering the test weight value
• Span calibration
• Optional five-point linearization
• Optional rezero calibration for test weights using hooks or chains
• Optional last zero calibration
• Optional temporary zero calibration

Note

The

880

requires the WZERO and WSPAN points to be calibrated. The linearity points are optional; they must

fall between zero and span, but must NOT duplicate zero or span. During calibration,

TARE

acts as a data

entry confirmation key. It also acts as an EXECUTE key, and accepts the value if calibration was successful.

SCALE

....

....

....

CALIBR

SETUP

....

....

Display and edit

test weight value

WVAL

WLIN

POINT1

Same as POINT 1

REZERO

Calibrating, please wait

Press Enter to

compensate for

zero offset

WGT 1

WZERO

Calibrating, please wait

Press Enter

to calibrate zero

Press Menu

to cancel

Previous A/D raw

counts are shown

CAL1

POINT2

POINT3

POINT4

POINT5

WSPAN

Calibrating, please wait

Press Enter

to calibrate span

Press Menu

to cancel

Previous A/D raw

counts are shown

LAST

Calibrating, please wait

Press enter to recall

the last established

pushbutton zero value

to allow calibration

without removing test

weights from the scale.

TEMP

Calibrating, please wait

Press enter to

temporarily zero the

displayed weight of a

non-empty scale.

After span calibration

the previously

calibrated zero value
is used as the offset.

Calibrating, please wait

Press Enter

to calibrate span

Press Menu

to cancel

Previous A/D raw

counts are shown

Display and edit

test weight value

See Section 4.2

See Section 4.3

Figure 4-1. Calibration (CALIBR) Menu