Rice Lake 880 Performance Series Indicator/Controller Technical/Service Manual User Manual

Page 52

Advertising
background image

Table 3-12. Ethernet Menu Parameters

ETHERNET Menu

Parameter

Choices

Description

DHCP

ON
OFF

Enables (ON) or disables (OFF) Dynamic Host Configuration Protocal.

IPADRS

000.000.000.000
Valid IP Address

Configures the IP address for this device.

NETMSK

000.000.000.000
Valid Net Mast

Specifies the subnet mask.

DFTGWY

000.000.000.000
Valid IP Address

Default Gateway.

DNSPRI

000.000.000.000
Valid IP Address

IP address for the primary DNS server.

DNSSEC

000.000.000.000
Valid IP Address

IP address for the secondary DNS server.

SERVER

See SERVER Sublevel below

CLIENT

See CLIENT Sublevel below

MAC

00.00.00.00.00.00

The MAC address for this device, read only.

SERVER/CLIENT Sublevel

RMOTIP

000.000.000.000
Valid IP Address

Remote IP address – IP address of the remote unit the 880 will connect to.

Client level only.

RMOTPT

1
1-65535

Remote port – TCP port number on the remote unit the 880 will connect to.

Client level only.

PORT

10001
1-65535

The TCP port number of the 880 server. Server level only.

TRIGGE

COMAND
STRIND
STRLFT

Selects the operation of the port.
•COMAND – allows operation of EDP commands and will print
•STRLFT – stream legal for trade scale data – data is transmitted at the configured

display update rate. Will also accept EDP commands.

•STRIND – stream industrial scale data – data is transmitted at the configured A/D

sample rate. Will also accept EDP commands.

TERMIN

CR/LF
CR

Termination. Selects the termination character(s) for data sent from the port.

EOLDLY

0
0-255

Port end-of-line delay, 0.1 second intervals.

TIMOUT

0
0-65535

Timeout – inactivity disconnect timeout. A connection (either client or server) is
dropped if there is no activity before the timeout expires. Time is in seconds. A
timeout value of 0 disables the inactivity disconnect.

SFMT

<2><P><W7.>

<U><M><S>

<CR><LF>

Stream format – specifies the stream format used for streaming output of scale data
(TRIGGE=STRLFT or STRIND). Alpha/numeric maximum length 200 characters.

ECHO

OFF
ON

Specifies whether characters received by the port are echoed back to the sending unit.

RESPNS

ON
OFF

Response – specifies whether the port transmits replies to serial commands.

Note

If an external device (such as a printer) that may transmit unexpected
data (such as a paper low message) is connected to the indicator the
response parameter should be set to OFF to prevent a reply from the
indicator from confusing the external device.

46

880 Technical/Service Manual