Setpoints

83

8.2

Batch Operations

Batches are controlled by digital inputs or EDP commands.

Batch Run

(BATRUN digital input) If a BATRUN digital input is configured, it must be active (low) for a batch to

be started, and for it to continue to run. If a batch is running and the input becomes inactive (high), it

will stop the batch at the current batch setpoint and turn off all associated digital outputs.

Batch Start

(BATSTR digital input or BATSTR EDP command) If the BATRUN digital input is active (low), or is

not assigned, batch start will start a batch, resume a paused batch or resume a stopped batch. If the

BATRUN digital input is inactive (high), batch start will reset the current batch.

Batch Pause

(BATPAS digital input or BATPAUSE EDP command) The BATPAS digital input will pause an

active batch, turning off all associated digital outputs EXCEPT those associated with CONCUR

and TIMER setpoints, while the input is active (low). As soon as the BATPAS digital input is made

inactive (high), the batch will resume.

BATPAUSE EDP command works the same, except the batch will not resume until a batch start

signal is received.

Batch Stop

(BATSTP digital input or BATSTOP EDP command) Stops an active batch at the current setpoint and

turns off all associated digital outputs.

Batch Reset

(BATRST digital input or BATRESET EDP command) Stops and resets an active batch to the

beginning of the process.

WARNING

To prevent personal injury and equipment damage, software-based interrupts must always be supplemented
by emergency stop switches and other safety devices necessary for the application.

Batching Switch

The batching switch option, PN 19369, comes as a complete unit in an FRP enclosure, with legend plate, locking

stop switch (mushroom button), and a run/start/abort 3-way switch.
Both switches are wired into the indicator’s digital I/O terminal strip as shown in Figure 8-1. Each switch uses a

separate digital input. Digital input #1 must be set to

BATSTR

and #2 must be set to

BATRUN

.

Once cables and switches have been connected to the indicator, use the setup switch to place the indicator in setup

mode. Use the DIG I/O menu (see Section 3.2.14 on page 54) to configure digital input and output functions.

RED

WHITE

BLACK

RED

BLACK

STOP/START MUSHROOM SWITCH

ABORT/RUN/START SWITCH

A
B
O
R
T

S
T
A
R
T

3

4

1

2

NO

NC

NC

NO

NO

+5VDC

GND

DIO1

DIO2

DIO3

DIO4

DIGITAL I/O

1

2

4

5

6

3

CPU BOARD

3

4

1

2

3

4

Figure 8-1. Batching Switch and Wiring Diagram Example