Rice Lake 880 Performance Series Indicator/Controller Technical/Service Manual User Manual

Page 44

Advertising
background image

Table 3-8. Feature Menu Parameters

FEATUR Menu

Parameter

Choices

Description

REGION

REGULA
REGWRD
DECFMT
TIME
DATE

Selects regional settings. See Level 3 sub menus.

UID

000000

Sets the unit ID, a string of up to 6 ASCII characters, which can be set via serial port or
keypad. This will be used in place of the <UID> token in a print format. The default value is “1.”

The Unit ID is also used as part of the File Name for configuration storage and printing to USB
Flash Drives.

CONSC#

CURVAL
RESVAL

Allows sequential numbering for print operations (CURVAL is the current value and RESVAL is
the reset value). The consecutive number value is incremented following each print operation
that includes <CN> in the ticket format. When the consecutive number is reset, it is reset to
the RESVAL specified in the parameter.

PASSWD

USER
SETUP

Sets a password to access the SETUP menu, or certain sub-menus in the USER menu.
Specify a non-zero value to enable the password.
• The SETUP password protects the entire SETUP menu, and when set is required even

when attempting entry into the SETUP menu using the setup switch.

• The USER password restricts access to the Time/Date, Accumulator, and Setpoints

sub-menus in the User menu.

Passwords can be overridden by loading new firmware, or entering 999999.

Note

Overriding passwords will clear configuration and calibration settings. To
preserve settings (i.e., ID information), use Revolution software to upload
the data to a PC, then download it back to the 880 after the password
override is performed.

KEYLCK

ZERO
GRSNET
UNITS
PRINT
TARE
MENU

Key Lock – Disables the listed keys. Select Lock to disable the key, and Unlock to enable the
key.

LOCALE

OFF
ON

Gravity compensation enable/disable.
Set this parameter ON to enable the LATUDE and ELEVAT

LATUDE

45
0-90

Press ENTER to display and edit the latitude in degrees for gravity adjustment to calibration.

(LOCALE must be set to ON)

ELEVAT

345
-9999 to +9999

Press ENTER to display and edit the elevation in meters for gravity adjustment to calibration.

(LOCALE must be set to ON)

38

880 Technical/Service Manual