Calibration

61

4.6

EDP Command Calibration

To calibrate the indicator using EDP commands, the indicator COM, USBCOM or Ethernet port must be connected

to a terminal or personal computer. See Section 2.5 on page 16 for cable connections.

Note

The indicator will respond with 'OK' if the value of the parameter was valid, or the command executed
properly. If the indicator responds with '??', then either the value for the parameter was invalid, or the
command could not be executed

Once the indicator is connected to the sending device, do the following:

1. Place the indicator in setup mode and remove all weight from the scale platform. If the test weights require

hooks or chains, place the hooks or chains on the scale for zero calibration.

2. Send the SC.WZERO#1 command to calibrate zero. The indicator displays

CALIBRATING, PLEASE WAIT

while calibration is in progress.

3. Place test weights on the scale and use the SC.WVAL#1 command to enter the test weight value in the

following format:

SC.WVAL#1=nnnnnn<CR>

4. Send the SC.WSPAN#1 command to calibrate span. The indicator displays

CALIBRATING, PLEASE WAIT

while calibration is in progress.

5. Up to five linearization points can be calibrated between the zero and span calibration values. Use the

following commands to set and calibrate a single linearization point:

SC.WLIN.V1#1=nnnnn<CR>
SC.WLIN.C1#1<CR>

The SC.WLIN.V1#1 command sets the test weight value (nnnnn) for linearization point 1. The

SC.WLIN.C1#1 command calibrates the point. Repeat using the SC.WLIN.Vn#1 and SC.WLIN.Cn#1

(where 'n' is the linearity point number) commands as required for additional linearization points.

6. To remove an offset value, clear all weight from the scale, including hooks or chains used to hang test

weights, then send the SC.REZERO#1 command. The indicator displays

CALIBRATING, PLEASE WAIT

while the zero and span calibrations are adjusted.

7. Send the KMENU or KEXIT EDP command to return to normal mode.

4.7

Revolution

®

Calibration

To calibrate the indicator using Revolution, the indicator serial port must be connected to a PC running the

Revolution configuration utility.

1.

Figure 4-2. Scale Button

Place the indicator in configuration mode (display reads

CONFIG

) and remove

all weight from the scale platform.

2. From Revolution, select

File

»

New

.

The Select Indicator dialog box appears.

3. Select 880 and click

OK

.

4. From the Communications menu, select

Connect

.

5. From the left pane, expand the Scale selection and select the

Scale

button.

6. From the Tools menu, select

Calibration Wizard

.

7. Select

NEXT

to begin the Calibration Wizard.

8. Select whether or not to perform a standard calibration or a standard with

multi-point linearization and select

NEXT

.

9. In the text box, enter the test weight value be to used for span calibration.
10. Select the check box if using chains or hooks during the calibration, then select

NEXT.

11. Remove all weight from the scale and select

Click to Calibrate Zero

to begin zero calibration. If the test

weights require an apparatus to hang, place it on the scale for zero calibration.

12. When zero calibration is complete, the Calibration Wizard prompts to place test weights on the scale. Place

the test weights on the scale, then select

Click to Calibrate Span

.

13. If choosing to perform linear calibration, the Calibration Wizard now displays prompts (1–5). Enter the

weight value for Linear Point #1, place test weights on scale and select

Measure

. Repeat for additional

linearization points, then select

NEXT

.

14. If the check box for using chains or hooks is selected, the Calibration Wizard prompts to perform a

Re-zero. Remove the apparatus used to hang the weights, and select

Click to Calibrate Re-Zero

.

15. The new and old calibration settings are displayed. To accept the new values, select

Finish

. To exit and

restore the old values, select

Cancel

.