26 RL-TS Series Operation Manual

14B

3.2 Counting Parts Count

The balance saves sample weight (unit weight) using the automatic memory update method (simplified SCS

method) to count the number of samples.
First, place a set number of samples. Next, place an appropriate number of additional samples, up to three

times the set number. Then, the balance will automatically update the average sample weight. Repeating

this step allows accurate counting.

1

Set to the function setting mode.
(Refer to “Section 2-5: Function Setting

Basics (P.

X

22

X

).)

Press and hold the [Function] key. After “Func” is

displayed, release the finger.

2

Select “Parts Counting.”

Press the [Zero/Tare] key several times to select “1.SEt

2

.”

3

Save the weighing mode.

Press the [Target] key.
The “parts counting” mode is set, displaying “Pcs.”

4

Start sampling.

<Press and hold>

Press and hold the [Function] key. After “U.SEt” is

displayed, release the finger.
The “on 10 Pcs” display indicates using ten samples.

Pressing the [Print] key during sampling can cancel the

sampling.

5

Select the number of samples.

Each press of the [Zero/Tare] key can select the sample

count between 5, 10, 30, and 100.
If the samples vary considerably in size or are lightweight,

set a greater number of samples.

6

Weigh the samples.

Place the set number of samples on the pan and then press

the [Function] key.
The displayed sample quantity (Example: “on 10 Pcs”)

blinks.”