6.1 Outputting to a Printer

By connecting a printer to the balance, span adjustment and span test results and measurement results can be
printed in ISO/GLP/GMP compliant form.

For printing examples, refer to “Appendix 3: Printing in Compliance with ISO/GLP/GMP”

(P.

X

91

X

).

58B

Connecting a printer

By using the D-SUB9P cable, connect the RS-232C connector of the balance to a printer.
The printers that can be connected to the balance are CSP-160 and CSP-240.
For the printer, the following setting is required. Set the following by referring to the operation manual of
the printer.
- Set the printing function (printing control) to “balance control.”

- Make the baud rate and other communication settings compatible with the settings made in the balance.

The date and time data set in the printer is also printed. Before printing, set the date and

time in the printer.

59B

Printing span adjustment and span test results

1

Set to the function setting mode.
(Refer to “Section 2-5: Function Setting

Basics (P.

X

22

X

).)

Press and hold the [Function] key. After

“Func” is displayed, release the finger.

2

Press the [Function] key several times to select

“E.GLP.”
Select “1” by pressing the [Zero/Tare] key.

3

Press the [Function] key several times to select

“E1.C.o.”
Select “1” by pressing the [Zero/Tare] key.

4

Save the setting.

Press the [Target] key.
The balance goes back to normal display.