80 RL-TS Series Operation Manual

Message

Cause

Remedy

The balance is influenced by static
electricity or noise.

- Unplug the AC adapter from the receptacle and then

turn the power on again.

- If this error occurs again, electric components may

have failed. Contact our Sales Office or Technical
Service Division.

The balance is influenced by static
electricity or noise.

- Unplug the AC adapter from the receptacle and then

turn the power on again.

- If this error occurs again, electric components may

have failed. Contact our Sales Office or Technical
Service Division.

The weight of a sample is too light at a
sampling during parts counting, or reference
weight saving during percentage weighing.

Use a heavier sample by referring to the Specifications
(P.

X

94

X

) to check the minimum unit weight and the

percentage weighing weight limit.

- At addition operation, you placed

additional samples on twice.

- At addition operation, you unloaded some

samples or you pressed the key without
adding samples.

- After setting the display to “0” (by unloading the

previous sample), place the next sample to continue
addition operation.

- Addition operation is impossible when 0 or a

negative value is displayed. Place a sample to
continue addition operation.

No inputs are sent from the weight sensor.

- Unplug the AC adapter from the receptacle and then

turn the power on again.

- If this error occurs again, the sensor may have failed.

Contact our Sales Office or Technical Service
Division.

Because the balance is unstable,
initialization cannot be completed.

The balance may be affected by an external influence
such as wind and vibration. Relocate the balance by
referring to the section “Do not use the balance in the
following places” (P. v).