76 RL-TS Series Operation Manual

- When either an O8 or O9 command is executed, it returns to the state of “O0”.

(3) Set measurement mode command

C1

C2

Code (C1)

Code (C2)

Description

Response

M

1

4DH

31H

Set to Mode 1

A00: Successful completion

E01: Command error

E02: Error

M

2

4DH

32H

Set to Mode 2

M

3

4DH

33H

Set to Mode 3

M

4

4DH

34H

Set to Mode 4

* The measurement mode to be activated by the above mode settings 1 to 4 depends on the weighing mode

currently in use.

- Relationship between weighing mode and mode setting

Mode

Weighing

machine

Parts

counting

Percentage

weighing

Unit

converting

Gravimeter

Animal

weighing

Mode 1

Weight
measuring

Weight
measuring

Weight
measuring

Weight
measuring

Error

Error

Mode 2 Gross weight

Parts
counting

Percentage
measuring

Coefficient
multiplying

Error

Error

Mode 3

Cumulative
weight*1

Cumulative
count*1

Cumulative
percent*1

Cumulative
sum*1

Error

Error

Mode 4

Display in
unit B*2

Average unit
weight

Error

Error

Error

Error

*1: Mode 3 (M3) can be specified only when the addition function is to be used. If the addition function is not

enabled, it results in an error.

*2: If a unit is NOT specified for unit B, the balance is set for the weight measuring mode. When you specify a

mode that is not supported by the current Weighing Mode, an error is returned from the balance.

(4)

Span adjustment/test command

C1

C2

Code (C1) Code (C2)

Description

Response

C

0

43H

30H

Disables command
inputs.*1

A00: Successful completion
E01: Command error
E02: Operation is disabled.
E03: Cancelled by operation
E04: Abnormal completion

C

3

43H

33H

Span adjustment with
external weight

C

4

43H

34H

Span test with external
weight

*1: A span adjustment and test commands will also be disabled.
* This command takes time because the balance sends back a response after an appropriate operation is

completed.

* If set to “[Cal] key disabled” (0) in the function setting “7.CA.,” a span adjustment and test command does

not work.

66B

Response

Upon receiving an input command, the balance sends out a response.

Advertising