56 RL-TS Series Operation Manual

5.2 Minimum Readability Setting

Use this function to set the minimum readability. The larger the minimum readability becomes, the less the
balance is affected by external influences. In addition, it takes less time for the balance reading to become
stable.
Each unit has different minimum readability.

1

Set to the function setting mode.
(Refer to “Section 2-5: Function Setting

Basics (P.

X

22

X

).)

Press and hold the [Function] key. After “Func” is

displayed, release the finger.

2

Select the minimum readability.

Press the [Function] key several times to select

“b2.d.A.”
Select from 1 through 5 by pressing the [Zero/Tare] key.

3

Save the setting.

Press the [Target] key.
The balance goes back to normal display.

To set the minimum readability of unit B, select “b4.d.b” in step 2.

You can also set the same unit for unit A and unit B, and set different minimum readability, so that unit
A and unit B can be used to switch the minimum readability.

Minimum readability example

Setting

value

TS-220~820

TS-2200~8200

TS-15K

g

kg

ct mom

g

kg

ct mom

g

kg

ct mom

1

0.01 0.00001 0.05 0.005 0.1

0.0001 0.5 0.05

1

0.001

5

0.5

2

0.02 0.00002 0.1 0.01

0.2

0.0002

1

0.1

2

0.002 10

1

3

0.05 0.00005 0.2 0.02

0.5

0.0005

2

0.2

5

0.005 10

2

4

0.1 0.0001 0.5 0.05

1

0.001

5

0.5

10

0.01

10

5

5

0.2 0.0002 1

0.1

2

0.002

10

1

10

0.02

10 10