26b5.4 power setting, 56bauto power off, 26b 5.4 – Rice Lake TS Series, Rice Lake Weighing Systems - Operation Manual User Manual

Page 62: Power setting, 56b auto power off, 4 power setting, Reference

Advertising
background image

58 RL-TS Series Operation Manual

26B

5.4 Power Setting

56B

Auto Power Off

This function, which is only available when the balance is operated by batteries, turns off the balance
automatically if left untouched for about five minutes.

1

Set to the function setting mode.

(Refer to “Section 2-5: Function
Setting Basics (P.

X

22

X

).)

Press and hold the [Function] key. After “Func” is
displayed, release the finger.

2

Set the auto power off.

Press the [Function] key several times to select
“9.A.P.”

Select “1” by pressing the [Zero/Tare] key.

3

Save the setting.

Press the [Target] key.

The balance goes back to normal display.

The auto power off function does not work under the following conditions:

- Function settings, time or date, or the interval function are set.
- An object is placed on the pan, and display is not stable.
- The balance is operated on the AC adapter.

To use the balance again after it is powered off, start from turning the power on.