25b5.3 saving container (tare) weight, 25b 5.3, Saving container (tare) weight – Rice Lake TS Series, Rice Lake Weighing Systems - Operation Manual User Manual

Page 61: 3 saving container (tare) weight

Advertising
background image

RL-TS Series Operation Manual 57

25B

5.3 Saving Container (Tare) Weight

Use this function to set a tare range when the balance is powered on using the latest saved tare weight. Use
this function when you turn the balance on or off with a sample and tare put on the pan.

Leaving the balance loaded with a substance and tare for a long period may result in a larger

error in weighing. You should set the tare range regularly.

1

Set to the function setting mode.
(Refer to “Section 2-5: Function Setting

Basics (P.

X

22

X

).)

Press and hold the [Function] key. After

“Func” is displayed, release the finger.

2

Set the tare saving.

Press the [Function] key several times to select

“J.tA.N.”
Select “1” by pressing the [Zero/Tare] key.

3

Save the setting.

Press the [Target] key.
The balance goes back to normal display.

4

Save the tare weight.

Place the container (tare) to be saved for weight

and then measure the tare.
The saved tare weight is updated every time a

tare range is set.

Advertising