62 RL-TS Series Operation Manual

28B

5.6 Improving the Stability of the Balance

When the balance is stable, “” is lit in the upper left of the display.
When displayed values flicker and stabilized display blinks, it indicates that the balance is influenced by
wind or vibration. In these situations, making a setting change can improve stability.
As greater values are set in the function setting of “stability judgment (4.5.d),” “response speed
(5.Re.),” and “minimum readability setting (b2.d.A, b4.d.b),” stability will be improved more.
Relationship between each function setting and wind/vibration influences

Wind/vibration

influences

Stability judgment

Response speed

Minimum readability

setting

Small

Big

1

1

1

2

2

2

3

3

3

4

4

4

5

5

In each of the functions, if wind and vibration influences are small, select 1 or 2. Set 3 to 5 for great
influences.

1

Set to the function setting mode.
(Refer to “Section 2-5: Function Setting

Basics (P.

X

22

X

).)

Press and hold the [Function] key. After “Func” is

displayed, release the finger.

2

Select each function.

Press the [Function] key several times to select the functions

(See the above table).
4.5.d. =

Stability judgment

5.r.E.=

Response speed

b2.d.A =

Minimum readability setting (A)

b4.d.b =

Minimum readability setting (B)

3

Select a setting.

Press the [Zero/Tare] key to select the setting values of each

function (see the above table).

Reference