RL-TS Series Operation Manual 47

55B

Judging by deviation values

The function setting for deviation value judgment is “24.Tp.2.”

75B

Judging by deviation values using the method of placing actual samples on the balance

Set a reference, upper limit, and lower limit values by weighing samples.

Check that the limit function is set for judging by deviation values (refer to the previous section “Limit
function setting”).

1

Set to the mode of using actual samples.

<Press and hold>

Press and hold the [Target] key. After “r.SEt”

is displayed, release the finger.

2

Weigh the reference weight sample.

Place the sample serving as the reference value

and then press the [Function] key.


3

Weigh the lower limit sample.

Place the sample serving as the lower limit and

then press the [Function] key.
For the case of one point setting, this operation

completes the setting saving. Go to step 5.
For the case of two point setting, “H.SEt” is

displayed.

4

Weigh the upper limit sample.

Place the sample serving as the upper limit

weight on the pan and then press the [Function]

key.
The judgment value is saved and then the

balance goes back to normal display.

5

Weigh.

Place a judgment target sample on the pan.

The OK, LO, or HI result judged from the upper and
lower limit values is displayed with the “◄” indication.