Rice Lake TS Series, Rice Lake Weighing Systems - Operation Manual User Manual

Page 79

Advertising
background image

RL-TS Series Operation Manual 75

Command format

(1) Tare range (zero-point adjustment) command

C1

C2

Code (C1) Code (C2)

Description

Value

Response

T

(SP) 54H

20H

- Tare range setting
- Zero-point

adjustment

None

A00: Successful completion
E01: Command error
E04: A tare range (zero-point

adjustment) cannot be set
(Range violation, weight error,
etc.).

(2) Set output control command

C1

C2

Code (C1) Code (C2)

Description

Response

O

0

4FH

-{}-30H

Stop output.

A00: Successful

completion

E01: Command

error

O

1

4FH

31H

Output continuously at all times.

O

2

4FH

32H

Output continuously if stable (Stop output if
unstable).

O

3

4FH

33H

Output once when the [Print] key is pressed
(whether the balance is stable or unstable).

O

4

4FH

34H

Output once when the balance is stable. Output
when a sample is lifted to cause the display to
indicate a value below zero, and then another
sample is placed to make the balance stable.

O

5

4FH

35H

Output once when the balance is stable. Stop
output when unstable. Output once when the
balance is stabilized again (the output includes
zero) even if it is not reloaded.

O

6

4FH

36H

Output once when the balance is stable. Output
continuously when unstable. Output is stopped
after a single output when the balance is stable
even if it is not reloaded.

O

7

4FH

37H

Output once when the [Print] key is pressed if
the balance is stable.

O

8

4FH

38H

Output once immediately.

O

9

4FH

39H

Output once after stabilized.

- Commands O0 to O7 have the same workings as the output control set by the function setting.
- Commands O8 and O9 are used to request data from the balance.
- Once executed, O0 to O7 commands are held. However, the status is reset to the function setting when the

balance is turned on again.