46 RL-TS Series Operation Manual

74B

Judging by absolute values using the method of entering values

Set upper and lower limits by entering values using the keys.

Check that the limit function is set for judging by absolute values (Refer to the previous section “Limit function
setting”).

1

Set to the mode of entering values.

<Press and hold>

Press and hold the [Target] key. After “L.SEt”

is displayed, release the finger.

2

Enter the lower limit value.

Enter the lower limit value with the following

steps:

1. Press the [Zero/Tare] key.

Zero blinks in the right end.

2. Select a number by pressing the

[Zero/Tare] key.

Pressing the key toggles the number

between 0 and 9, and decimal point.

3. Pressing the [Function] key shifts the

blinking digit to the left, and the next

lower-order digit blinks.

4. Press the [Target] key to save the value.


For the case of one point setting, this operation

completes the setting saving. Go to step 4.
For the case of two point setting, “H.SEt” is

displayed.

3

Enter the upper limit value.

Enter the upper limit value by using the same

operation as with step 2.
The judgment value is saved and then the

balance goes back to normal display.

4

Weigh.

Place a judgment target sample on the pan.
The OK, LO, or HI result judged from the upper

and lower limit values is displayed with the “◄”

indication.