57bauto backlight off, 27b 5.5 id no. setting, 57b auto backlight off – Rice Lake TS Series, Rice Lake Weighing Systems - Operation Manual User Manual

Page 63: 27b 5.5, Id no. setting, Reference, Auto backlight off

Advertising
background image

RL-TS Series Operation Manual 59

57B

Auto Backlight Off

This function automatically turns off the backlight if the balance is left untouched in measurement mode for about
three minutes.

1

Set to the function setting mode.
(Refer to “Section 2-5: Function Setting

Basics (P.

X

22

X

).)

Press and hold the [Function] key. After “Func” is

displayed, release the finger.

2

Set the auto backlight off.

Press the [Function] key several times to select “A.A.b.”
Select “1” by pressing the [Zero/Tare] key.

3

Save the setting.

Press the [Target] key.
The balance goes back to normal display.

The auto backlight off function does not work under the following conditions:

- Function settings are set.
- An object is placed on the pan, and display is not stable.

Placing an object on the pan or pressing any key turns the backlight on again automatically.